Cardionetics C.Net5000
Technical Specification
| Electrode Pad Placement | Pad positions: RA, LA, V4, Lead: chest-modified V4 |
| Heart Rate | Mean (4 intervals)
Minimum: 20 bpm Maximum: 250 bpm Tachycardia: over 120 bpm, minimum 30 seconds Bradycardia: under 50 bpm, minimum 30 seconds |
| Wide Complex Tachycardia | Heart rate: 120 bpm to 300 bpm
Duration: minimum 3 beats Morphology: wide, unifocal |
| ST Segment Analysis | Depression: 1 mm (2 mm if upsloping)
Measured point: J + 60 ms (J + 50 ms in tachycardia) Isoelectric point: PR segment Slope: upsloping, horizontal, or downsloping Duration: minimum 30 seconds |
| AF Detection | Irregular irregular RR intervals
Variable P wave morphology Duration: minimum 30 seconds |
| Asystole | Pause: 1.7 to 3.0 seconds Arrest: over 3.0 seconds Missed beats: pauses in synchronous rhythm |
| Ectopic Beats | Ventricular: unifocal, multifocal
Atrial: premature (less than 75% average RR interval) Classification: isolated, trigeminy, bigeminy, couplet, triplet, salvo (4 to 7 beats), episode (8 or more beats) |
| Symptom Button | Up to 10 events |
| Display | Monochrome LCD with backlight
Resolution: 128 × 64 pixels Sweep speed: 12.5 mm/second Gain: 5 mm/mV |
| Report Printout | ECG trace: 24 seconds, 25 mm/second
Trend graph: 24 hours, 25 mm/hour |
| Physical | Size: 107 mm × 80 mm × 27 mm
Weight: 200 g Power: 2 × AA batteries Index of protection: IP20 rated by design |
| Electrical | Channels: single bipolar
Sample frequency: 100 Hz Sample resolution: 12-bit Frequency response: 0.1 Hz to 23 Hz Input impedance: greater than 20 MΩ |
| Environmental | Operating temperature: 0°C to 50°C
Storage temperature: 0°C to 60°C Relative humidity: 20% to 65% noncondensing Shock and vibration: AAMI EC38, 4.2.1.2 |
| Regulatory Standards | CE mark: MDD, 93/42/EEC
Performance: ANSI/AAMI EC38 Validation: ANSI/AAMI EC57 Quality: BS EN ISO 13485:2003 Safety: BS EN 60601-1 (1997) |
|
|